he MA24350A power sensor is designed to provide fast, accurate average power measurements from 10 MHz to 50 GHz over 90 dB of dynamic range. The sensor employs high-performance digital processing that enables measurements speeds of >2,100 continuous power readings/s and >5,600 buffered readings/s.
A unique, low noise design eliminates the need to zero the sensor before taking measurements for most applications.
The sensors have internal and external triggering capability that facilitates time-based measurements and the use of list mode to speed up automated processes.
The sensor can be controlled with a PC via remote programming commands or with PowerXpert™, a free of charge software application. These sensors are also compatible with most Anritsu RF and microwave handheld instruments. The high accuracy power meter software option (e.g., Option 19) may be required.
The excellent damage protection combined with the silicone protective covering give the sensor increased ruggedness and enable users to protect their investment.
Key Features |
Frequency range: 10 MHz — 50 GHz |
Power measurement range: +20 dBm — -70 dBm |
CW average power measurements |
Fast measurement speed: >2,1001 readings/s continuous, >5,600 readings/s buffered¹ |
Able to accept high power levels before being damaged: +26 dBm CW, +32 dBm 10 µs |
No zero required* |
NIST traceable calibrations |
Compatible with Anritsu handheld instruments |
Silicone protective covering for additional field durability |
Worldwide calibration and service centers |
Note:
¹ Speed may vary depending on the speed of the CPU controlling the sensor
* for signals > -45 dBm
Specifications |
|
MA24350A
Microwave CW USB Power Sensor
|
|
Feature | Parameter |
Frequency Range | 10 MHz — 50 GHz, V(m) Connector |
Dynamic Range | –70 dBm — +20 dBm |
SWR (x:1) | <1.90, < 50 MHz
<1.17, 50 MHz — 150 MHz
<1.08, 150 MHz — 2 GHz
<1.16, 2 GHz — 6 GHz
<1.21, 6 GHz — 18 GHz
<1.29, 18 GHz — 33 GHz
<1.44, 33 GHz — 40 GHz
|
Measurement Uncertainty¹
(dB)
| +20 dBm — +15 dBm / +15 dBm — -70 dBm
0.18 0.15 10 MHz — 18 GHz
0.21 0.17 18 GHz — 50 GHz
|
RF Connector | V(m) Connector |
Note:
¹ Expanded uncertainty with K=2 for CW measurements after zero operation.
Includes linearity over temperature uncertainties, but not the effects of mismatch, zero set and drift, or noise.
Specifications apply over a temperature range of 0 °C — 50 °C and a 60 minute warm up period.
Specifications |
|
MA243xxA
Microwave CW USB Power Sensor
|
|
Model Overview |
|
Model | Description |
MA24330A | 33 GHz USB Power Sensor |
MA24340A | 40 GHz USB Power Sensor |
MA24350A | 50 GHz USB Power Sensor |
Included Accessories |
|
10585-00021 | Quick Start Guide |
2000-1605-R | 1.5 m BNC(m) to MCX(m) cable |
2000-2010-R | 1.83 m USB A to Micro-B cable |
Options |
|
MA24330A-097 | Option 97:
ISO/IEC 17025 and ANSI/NCSL Z540-1 or ANSI/NCSLI Z540.3
(includes test report, uncertainty data, and accreditation symbol)
|
MA24330A-098 | Option 98:
Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
|
MA24330A-099 | Option 99:
Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
(includes test report and uncertainty data)
|
MA24340A-097 | Option 97:
ISO/IEC 17025 and ANSI/NCSL Z540-1 or ANSI/NCSLI Z540.3
(includes test report, uncertainty data, and accreditation symbol)
|
MA24340A-098 | Option 98:
Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
|
MA24340A-099 | Option 99:
Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
(includes test report and uncertainty data)
|
MA24350A-098 | Option 98:
Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
|
MA24350A-099 | Option 99:
Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
(includes test report and uncertainty data)
|