配置型号包含以下特性:
MultimeterData AcquisitionCounters & Function GeneratorsDigital Multimeters请参考资源中的资料文档或查看产品总览。
如果您想了解更多,请 联系,我们的团队将为您提供帮助。
The Fluke 8845A 6.5 digit precision multimeter handles your most demanding measurements on the bench or in a system. It measures volts, ohms, and amps, with basic V dc accuracy of up to 0.0024%, 10 A current range, and a wide ohms range. The 8845A also measures frequency and period. The functions of a counter, capacitance meter, and thermometer are built in. Graphical display modes include Trendplot paperless recorder mode, statistics and hist histograms.
| Key Features |
| 6.5 digit resolution |
| Basic V DC accuracy of up to 0.0035% |
| Dual display |
| 100 μA to 10 A current range, with up to 100 pA resolution |
| Wide ohms range from 100 Ω to 100 MΩ with up to 100 μΩ resolution |
| 2 x 4 ohms 4-wire measurement technique |
| Measure frequency and period |
| Fluke 45 and Agilent 34401A emulation |
| Graphical display |
| Trendplot™ paperless recorder mode, statistics, histogram |
| CAT I 1000 V, CAT II 600 V |
| Benefits | |
| Dual Display and Graphical Capabilities | A dual display lets you measure two parameters of the same signal from one test connection. Use the unique analyze mode to view data as a real time trend plot or histogram. |
| Perform 4-Wre Measurement with only Two Leads | Patented split terminal jacks for the 2 x 4 ohms function let you perform 4-wire measurements using only two leads instead of four. |
| Multiple Connectors give you Maximum Flexibility | Serial, IEEE-488, and Ethernet interfaces come standard. |
| FlukeView Software and Drivers add PC Power and Application Compatibility | FlukeView Forms lets you document, store and analyze individual readings or series of measurements, then convert them into professional-looking documents. Instrument programming drivers, DLL files and support information are included to get your test system up and running as quickly as possible |
| FlukeView Forms Basic | The Fluke 8845A ships with a free copy of FlukeView Basic. For a larger array of forms or to use FlukeView Forms Designer to customize your forms, upgrade to FlukeView Forms with FVF-UGg> |
| Applications |
| Manufacturing Test |
| Research and Development |
| Services |
| Specifications | ||
Fluke 884xA 6.5 Digit Precision Digital Multimeter | ||
| Model Overview | ||
| Model | 8845A | 8846A |
| Display | VFD Dot Matrix | VFD Dot Matrix |
| Resolution | 6.5 Digits | 6.5 Digits |
| V DC | ||
| Ranges | 100 mV to 1000 V | 100 mV to 1000 V |
| Max. Resolution | 100 nV | 100 nV |
| Accuracy | 0.0035 + 0.0005 | 0.0024 + 0.0005 |
| V AC | ||
| Ranges | 100 mV to 750 V | 100 mV to 1000 V |
| Max. Resolution | 100 nV | 100 nV |
| Accuracy | 0.06 + 0.03 | 0.06 + 0.03 |
| Frequency | 3 Hz to 300 KHz | 3 Hz to 300 KHz |
| Resistance | ||
| 2x4 Wire | Yes | Yes |
| Ranges | 100Ù to 100 MÙ | 10Ù to 1 GÙ |
| Max. Resolution | 100 µÙ | 10 µÙ |
| Accuracy | 0.010 + 0.001 | 0.010 + 0.001 |
| A DC | ||
| Ranges | 100 µA to 10 A | 100 µA to 10 A |
| Max. Resolution | 100 pA | 100 pA |
| Accuracy | 0.050 + 0.005 | 0.050 + 0.005 |
| A AC | ||
| Ranges | 10 mA to 10 A | 100 µA to 10 A |
| Max. Resolution | 10 µA | 100 pA |
| Accuracy | 0.10 + 0.04 | 0.10 + 0.04 |
| Frequency | 3 Hz to 10 kHz | 3 Hz to 10 kHz |
| Freq/Period | ||
| Ranges | 3 Hz to 300 kHz | 3 Hz to 1 MHz |
| Max. Resolution | 1 µHz | 1 µHz |
| Accuracy | 0.01% | 0.01% |
| Continuity/ Diode Test | Yes | Yes |
| Capacitance | ||
| Ranges | — | 1 nF to 0.1 F |
| Max. Resolution | — | 1 pf |
| Accuracy | — | 1% |
| Temperature | ||
| Type | Platinum RTD | |
| Range | — | -200 °C to +600 °C |
| Max. Resolution | — | 0.01 ° |
| Accuracy | — | 0.06 ° |
| Math Functions | Zero, Min/Max/Average, Std Dev; mx+b | Zero, Min/Max/Average, Std Dev; mx+b |
| dB/dBm | Yes | Yes |
| Advanced Functions | ||
| Statistics/Histogram | Yes | Yes |
| Trend plot | Yes | Yes |
| Limit Test | Yes | Yes |
| Input Output | ||
| USB Memory | — | USB memory drive port |
| Real Time Clock | — | Yes |
| Interfaces | RS 232 IEE-488.2 Ethernet USB (with optional adaptor) | RS 232 IEE-488.2 Ethernet USB (with optional adaptor) |
| Programming Languages/ Modes | SCPI (IEEE-488.2) Agilent 34401A Fluke 45 | SCPI (IEEE-488.2) Agilent 34401A Fluke 45 |
| Specifications | ||
Fluke 884xA 6.5 Digit Precision Digital Multimeter | ||
| Model Overview | ||
| Model | 8845A | 8846A |
| Weight | 3.6 kg (8.0 lbs) | 3.6 kg (8.0 lbs) |
Size (HxWxD) | 88 mm x 217 mm x 297 mm (3.46 in x 8.56 in x 11.7 in) | 88 mm x 217 mm x 297 mm (3.46 in x 8.56 in x 11.7 in) |
| Safety | Designed to comply with IEC: 61010-1 2000-1 ANSI/ISA-S82.01-1994 CAN/CSA-C22.2 No. 1010.1-92 CAT I 1000 V CAT II 600 V | Designed to comply with IEC: 61010-1 2000-1 ANSI/ISA-S82.01-1994 CAN/CSA-C22.2 No. 1010.1-92 CAT I 1000 V CAT II 600 V |
| Warranty | One year | One year |
| Specifications | |
Fluke 884xA 6.5 Digit Precision Digital Multimeter | |
| Model Overview | |
| Model | Description |
| 8845A | 6.5 digit precision multimeter, 35 ppm |
| 8846A | 6.5 digit precision multimeter, 24 ppm USB mem |
| Options and Accessories | |
| TL71 | Test Lead Set, Premium DMM |
| TL910 | Precision Electronic Probe Set |
| 884X-SHORT | 4-Wire Short |
| TL2X4W-PT | 2x4 Wire Ohms Test Lead 2 mm Probe Tip |
| 884X-RTD | 100 Ohm RTD Temperature Probe |
| 884X-512M | USB Memory 512 M |
| 884X-1G | USB Memory 1 GB |
| Y8022 | IEEE488 cable (2 m) |
| 884X-USB | USB to RS232 cable adapter |
| FVF-UG | FlukeView Forms Software Upgrade—No Cable |
| Y8846 | Rack Mount Kit |